Skip Navigation

Publication Detail

Title: Deployment, Calibration, and Cross-Validation of Low-Cost Electrochemical Sensors for Carbon Monoxide, Nitrogen Oxides, and Ozone for an Epidemiological Study.

Authors: Zuidema, Christopher; Schumacher, Cooper S; Austin, Elena; Carvlin, Graeme; Larson, Timothy V; Spalt, Elizabeth W; Zusman, Marina; Gassett, Amanda J; Seto, Edmund; Kaufman, Joel D; Sheppard, Lianne

Published In Sensors (Basel), (2021 Jun 19)

Abstract: We designed and built a network of monitors for ambient air pollution equipped with low-cost gas sensors to be used to supplement regulatory agency monitoring for exposure assessment within a large epidemiological study. This paper describes the development of a series of hourly and daily field calibration models for Alphasense sensors for carbon monoxide (CO; CO-B4), nitric oxide (NO; NO-B4), nitrogen dioxide (NO2; NO2-B43F), and oxidizing gases (OX-B431)-which refers to ozone (O3) and NO2. The monitor network was deployed in the Puget Sound region of Washington, USA, from May 2017 to March 2019. Monitors were rotated throughout the region, including at two Puget Sound Clean Air Agency monitoring sites for calibration purposes, and over 100 residences, including the homes of epidemiological study participants, with the goal of improving long-term pollutant exposure predictions at participant locations. Calibration models improved when accounting for individual sensor performance, ambient temperature and humidity, and concentrations of co-pollutants as measured by other low-cost sensors in the monitors. Predictions from the final daily models for CO and NO performed the best considering agreement with regulatory monitors in cross-validated root-mean-square error (RMSE) and R2 measures (CO: RMSE = 18 ppb, R2 = 0.97; NO: RMSE = 2 ppb, R2 = 0.97). Performance measures for NO2 and O3 were somewhat lower (NO2: RMSE = 3 ppb, R2 = 0.79; O3: RMSE = 4 ppb, R2 = 0.81). These high levels of calibration performance add confidence that low-cost sensor measurements collected at the homes of epidemiological study participants can be integrated into spatiotemporal models of pollutant concentrations, improving exposure assessment for epidemiological inference.

PubMed ID: 34205429 Exiting the NIEHS site

MeSH Terms: Air Pollutants*/analysis; Air Pollution*/analysis; Calibration; Carbon Monoxide/analysis; Environmental Monitoring; Epidemiologic Studies; Humans; Nitric Oxide/analysis; Nitrogen Dioxide/analysis; Ozone*/analysis; Particulate Matter/analysis

Back
to Top