Skip Navigation

Publication Detail

Title: A novel device for measuring respirable dustiness using low-mass powder samples.

Authors: O'Shaughnessy, Patrick T; Kang, Mitchell; Ellickson, Daniel

Published In J Occup Environ Hyg, (2012)

Abstract: Respirable dustiness represents the tendency of a powder to generate respirable airborne dust during handling and therefore indicates the propensity for a powder to become an inhalation hazard. The dustiness of 14 powders, including 10 different nanopowders, was evaluated with the use of a novel low-mass dustiness tester designed to minimize the use of the test powder. The aerosol created from 15-mg powder samples falling down a tube were measured with an aerodynamic particle sizer (APS). Particle counts integrated throughout the pulse of aerosol created by the falling powder were used to calculate a respirable dustiness mass fraction (D, mg/kg). An amorphous silicon dioxide nanopowder produced a respirable D of 121.4 mg/kg, which was significantly higher than all other powders (p < 0.001). Many nanopowders produced D values that were not significantly different from large-particle powders, such as Arizona Road Dust and bentonite clay. In general, fibrous nanopowders and powders with primary particles >100 nm are not as dusty as those containing granular, nano-sized primary particles. The method used here, incorporating an APS, represents a deviation from a standard method but resulted in dustiness values comparable to other standard methods.

PubMed ID: 22335240 Exiting the NIEHS site

MeSH Terms: No MeSH terms associated with this publication