Skip Navigation

Publication Detail

Title: Rapid Analysis of the Size Distribution of Metal-Containing Aerosol.

Authors: Park, Jae Hong; Mudunkotuwa, Imali A; Crawford, Kathryn J; Anthony, T Renée; Grassian, Vicki H; Peters, Thomas M

Published In Aerosol Sci Technol, (2017)

Abstract: Conventional methods to measure the metallic content of particles by size are time consuming and expensive, requiring collection of particles with a cascade impactor and subsequent metals analysis by inductively coupled plasma mass spectrometry (ICP-MS). In this work, we describe a rapid way to measure the size distribution of metal-containing particles from 10 nm to 20 μm, using a nano micro-orifice uniform-deposit impactor (nano-MOUDI) to size-selective and collect particles that are then analyzed with a field portable X-ray fluorescence (FP-XRF) to determine metal composition and concentration. The nano-MOUDI was used to sample a stainless-steel aerosol produced by a spark discharge system. The particle-laden substrates were then analyzed directly with FP-XRF and then with ICP-MS. Results from FP-XRF were linearly correlated with results from ICP-MS (R2 = 0.91 for Fe and R2 = 0.84 for Cr). Although the FP-XRF was unable to detect Fe particles at mass per substrate loadings less than 2.5 μg effectively, it produced results similar to those using the ICP-MS at a mass per substrate loading greater than 2.5 μg.

PubMed ID: 28871214 Exiting the NIEHS site

MeSH Terms: No MeSH terms associated with this publication

Back
to Top